Interior Structure built in steel pipe with coating of epoxy paint specially resistant to acids, bases and alkalis.
Working surfaces: according to applications)
• Continuous acid-resisting tiling of 30 mm. in thickness, with peripheral rim to protect against spills of acids and bases.
• Stainless steel.
• Polypropylene.
• Compacted resin.
• PVDF.
Low piece constituted by two cupboards with interior panel, easily removable for possible inspections.
• Optionally special piece for acids and bases with possibility of gas extraction.
Front window sliding vertically, mounted on aluminium frame and sliding in guide rails.
• Safety glass of 6 mm., slides transversely to allow lateral opening of the area for protection of the operator.
• System of counterweights and pulleys with rollers for maximum opening of the window or in two stages, the latter being adjustable in height.
• Base of the frame, formed by an aerodynamic rail in polypropylene that allows air to enter at the bottom while protecting against splashing at the same time.
• Obstacle detector in the closing area.
Uniform distribution of air along the whole cabinet with free evacuation of turbulence, with vertical - lateral inlets, at top and bottom.
Interior baffle of 6 mm. in thickness formed by two horizontal panels for suction of heavy gases.
Extractor motor with IP55 protection.
• Power appropriate for the cabinet type and characteristics of the building.
• Possibility of fitting in the cabinet or in the roof.
Accesories:
• Exhaust diffuser in PPs.
• Base in compacted resin with anti-vibration mechanisms.
• Structure for adjustable base in galvanized steel.
• Protective casing in PE.
Water service formed by tap with remote control and tiled basin with drainage and siphon, piping for supply of cold water.
Gas service formed by tap with remote control, copper pipe for gas supply.
Lighting service fluorescent tube of 40W with reflective body in steel plate protected with acrylic glass of prismatic structure, serving in turn as a safety ceiling in the event of explosion inside the cabinet. Secured against falling.
Electrical service and controls: Located in the left column of the cabinet, incorporating:
• 4 electric sockets 230V., 16A.
• 1 switch for fluorescent tube 230V., 16A.
• 1 magneto-thermal breaker.
• 1 motor guard for extractor.
• 1 controller of the air extraction flow.
Air flow controller:
Electronic system for control of the extractor operation in compliance with the DIN 12924 standard.
• Accumulator for operation with power cut during 24 hours, for system failure warning.
• Vetilation of bottom cupboard.
• Temperature control with alarms.
.: Optionally TECLAB®
• Regulation of air flow to constant speed.
• Power-operated window closing actuated on detecting presence.
• Scrubbing of gases by irrigation.
• Automatic extinguishing of fires.
• Safety panel with 2 openings for interior manipulation.
• Lower safety cupboard for acids and bases.
.: Special Cabinets TECLAB®
• Cabinets for radio isotopes.
• Cabinets with water irrigation system for scrubbing gases (perchloric acid).
• Cabinets for hydrofluoric acid.
• Cabinets for concentrated acids..